dc.description.abstractA micro-spectrometer has been realized based on an array of Fabry-Perot thin-film optical resonators. The 16 channel micro-spectrometer is IC fabrication compatible and operates in the visible spectral range with an inter-channel shift of 6 nm. Each of the channels is sensitive in a single peak with FWHM of 16 nm. A FWHM < 2 nm and finesse of 40 for narrowband operation is demonstrated.eng
dc.description.sponsorshipSTW - Project DEL 55.3733. TUDelft. Fundação para a Ciência e Tecnologia - Praxis XXI-BD/5181/95.eng
dc.language.isoengeng
dc.rightsopenAccesseng
dc.subjectOptical filterseng
dc.subjectVisible rangeeng
dc.titleHigh-selectivity single-chip spectrometer for operation at visible wavelengthseng
dc.typeconferencePapereng
dc.peerreviewedyeseng
Aparece nas coleções:DEI - Artigos em atas de congressos internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
IEDM98.pdfDocumento principal113,82 kBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID