Title: | Biological microsystem for measuring uric acid in biological fluids |
Author(s): | Minas, Graça Martins, Júlio S. Ribeiro, J. C. Wolffenbuttel, R. F. Correia, J. H. |
Keywords: | Bio-system Uric acid Optical absorption |
Issue date: | 2004 |
Publisher: | Elsevier |
Journal: | Sensors and Actuators A: Physical |
Citation: | "Sensors and Actuators A : Physical". ISSN 0924-4247. 110 (2004) 33-38. |
Abstract(s): | This paper describes a biological microsystem (bio-system) for measuring uric acid concentration in serum, plasma or urine. Its operation is based on optical absorption in a well-defined part of the visible spectrum. The bio-system is composed by two dies: one is fabricated in polystyrene and contains the microchannels and the other is fabricated in a CMOS standard process and contains the photodetector and readout electronics. The uric acid concentration is measured by using a mixture of 14µl of infinity™ uric acid reagent with 0.25µl of sample. The achieved sensitivity is 0.33 mg/dl (±0.6% of the value in urine of a healthy person), with a 1mm lightpath. Using an optical absorption method, a maximum peak at wavelength λ = 494 nm, is detected. This bio-system can be included in the group of low-cost disposable devices for biological fluids analysis. |
Type: | Article |
URI: | https://hdl.handle.net/1822/1986 |
DOI: | 10.1016/j.sna.2003.10.049 |
ISSN: | 0924-4247 |
Peer-Reviewed: | yes |
Access: | Open access |
Appears in Collections: | DEI - Artigos em revistas internacionais |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
science.pdf | 235,63 kB | Adobe PDF | View/Open |