Utilize este identificador para referenciar este registo: https://hdl.handle.net/1822/73945

TítuloA preliminary analysis of the wear pathways of sliding contacts on temporomandibular joint total joint replacement prostheses
Autor(es)Pinto-Borges, Henrique
Carvalho, Óscar Samuel Novais
Henriques, Bruno Alexandre Pacheco de Castro
Silva, Filipe Samuel
Ramos, António
Souza, Júlio C. M.
Palavras-chaveDebris
Temporomandibular joint total joint replacement
Wear
Toxicity
Data22-Abr-2021
EditoraMultidisciplinary Digital Publishing Institute (MDPI)
RevistaMetals
CitaçãoPinto-Borges, H.; Carvalho, O.; Henriques, B.; Silva, F.; Ramos, A.; Souza, J.C.M. A Preliminary Analysis of the Wear Pathways of Sliding Contacts on Temporomandibular Joint Total Joint Replacement Prostheses. Metals 2021, 11, 685. https://doi.org/10.3390/met11050685
Resumo(s)In the last years, several materials and design have been assessed in an attempt to improve the mechanical performance of temporomandibular joint total joint replacement (TMJ TJR) prostheses. However, the wear of the TMJ TJR condyle to the polymer-based fossa component during loading and sliding movements. That promotes the release of debris and risks of toxicity to the surrounding tissues. The purpose of this study was to perform a narrative literature review on the wear of TMJ TJR sliding contacts and potential toxicity of metallic debris to the patients. Previous studies reported a significant deterioration of the sliding contact surfaces of TMJ TJR prostheses. Material loss as a result of wear can cause a TMJ TJR condyle/fossa mismatch and the modification of the contact pressure and chewing loading. As a further consequence of wear, metal particles are released to the surrounding tissues with a high risk of local tissue and systemic toxicity through the bloodstream. The presence of particles induces the stimulation of inflammatory reactions depending on the concentration and size of debris. Thus, CoCr-based condyle release metallic ions and sub-micron particles that can be engulfed by macrophages or internalized by other tissue cells. The wear and material loss of TMJ TJR could be decreased by design optimization and novel materials with low friction and contact pressure. That consequently decrease the amount of metallic ions and particles to the surrounding tissues, preventing peri-prosthetic inflammatory reactions.
TipoArtigo
URIhttps://hdl.handle.net/1822/73945
DOI10.3390/met11050685
e-ISSN2075-4701
Versão da editorahttps://www.mdpi.com/2075-4701/11/5/685
Arbitragem científicayes
AcessoAcesso aberto
Aparece nas coleções:BUM - MDPI

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
metals-11-00685.pdf1,67 MBAdobe PDFVer/Abrir

Este trabalho está licenciado sob uma Licença Creative Commons Creative Commons

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID