Utilize este identificador para referenciar este registo: https://hdl.handle.net/1822/53325

TítuloThe feasibility of a custom-made endoprosthesis in mandibular reconstruction: implant design and finite element analysis
Autor(es)Pinheiro, Manuel
Alves, J. L.
Palavras-chaveBiomechanical Phenomena
Bone-Implant Interface
Computer Simulation
Dental Implants
Dental Stress Analysis
Humans
Mandible
Mandibular Reconstruction
Stress, Mechanical
Finite Element Analysis
Mandibular Prosthesis
Cephalometry
Endosseous implantation
Mandibular prosthesis implantation
DataDez-2015
EditoraElsevier 1
RevistaJournal of Cranio-Maxillofacial Surgery
Resumo(s)This work studies the feasibility of custom-made endoprosthesis in the reconstruction of major mandibular defects. The natural anatomical and occlusal relations are used to accurately reconstruct a mandibular defect. The customized implant allows the accurate restoration of the facial profile and aesthetics. The biomechanical behaviour of mandibular endoprosthesis was validated with Finite Element Analysis for three masticatory tasks, namely incisal, right molar and left group clenching. The implanted mandible shows displacement fields and stress distributions very similar to the intact mandible. The strain fields observed along the bone-implant interface may promote bone maintenance and ingrowth. The preliminary results show that this implant may be a reliable alternative to other prosthetic mandibular reconstruction approaches.
TipoArtigo
URIhttps://hdl.handle.net/1822/53325
DOI10.1016/j.jcms.2015.10.004
ISSN1010-5182
Arbitragem científicayes
AcessoAcesso aberto
Aparece nas coleções:DEM - Artigos em revistas de circulação internacional com arbitragem científica

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
art16.pdf3,72 MBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID