TítuloCentral corneal thickness and anterior chamber depth measurement by Sirius® Scheimpfug tomography and ultrasound
Autor(es)Jorge, Jorge
Rosado, J. L.
Diaz-Rey, José Alberto
González-Méijome, José Manuel
Palavras-chaveScheimpflug corneal tomography
Ultrasound biometry
Ultrasound pachymetry
Limits of agreement
Scheimpfug corneal tomography
Data2013
EditoraDove Press Ltd
RevistaClinical Ophthalmology
Resumo(s)Background: The purpose of this study was to compare the accuracy of the new Sirius® Scheimpflug anterior segment examination device for measurement of central corneal thickness (CCT) and anterior chamber depth (ACD) with that of CCT measurements obtained by ultrasound pachymetry and ACD measurements obtained by ultrasound biometry, respectively. Methods: CCT and ACD was measured in 50 right eyes from 50 healthy subjects using a Sirius Scheimpflug camera, SP100 ultrasound pachymetry, and US800 ultrasound biometry. Results: CCT measured with the Sirius was 546 ± 39 μm and 541 ± 35 μm with SP100 ultrasound pachymetry (P = 0.003). The difference was statistically significant (mean difference 4.68 ± 10.5 μm; limits of agreement −15.8 to 25.20 μm). ACD measured with the Sirius was 2.96 ± 0.3 mm compared with 3.36 ± 0.29 mm using US800 ultrasound biometry (P , 0.001). The difference was statistically significant (mean difference -0.40 ± 0.16 mm; limits of agreement -0.72 to 0.07 mm). When the ACD values obtained using ultrasound biometry were corrected according to the values for CCT measured by ultrasound, the agreement increased significantly between both technologies for ACD measurements (mean difference 0.15 ± 0.16 mm; limits of agreement -0.16 to 0.45 mm). Conclusion: CCT and ACD measured by Sirius and ultrasound methods showing good agreement between repeated measurements obtained in the same subjects (repeatability) with either instrument. However, CCT and ACD values, even after correcting ultrasound ACD by subtracting the CCT value obtained with either technology should not be used interchangeably. Keywords: Scheimpflug corneal tomography, ultrasound biometry, ultrasound pachymetry, limits of agreement.
TipoArtigo
URIhttps://hdl.handle.net/1822/27020
DOI10.2147/opth.S35121
ISSN1177-5483
Arbitragem científicayes
AcessoAcesso aberto
Aparece nas coleções:CDF - OCV - Artigos/Papers (with refereeing)

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
5.6.38. JORGE_12_CLIN OPHTHALMO (Sirius Topographer).pdf1,03 MBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID