TitleIntraocular pressure measurement with ocular response analyzer over soft contact lens
Author(s)Sapkota, Kishor
Franco, Sandra
Lira, Madalena
KeywordsOcular response analyzer
Corneal compensated intraocular pressure
Goldmann correlated intraocular pressur
Soft contact lens
Goldmann correlated intraocular pressure
Issue dateDec-2014
PublisherElsevier
JournalContact Lens and Anterior Eye
Abstract(s)Purpose: To compare intraocular pressure (IOP) measured with ocular response analyzer (ORA) with and without soft contact lenses (CL) on eye. Methods: Goldmann correlated intraocular pressure (IOPg) and corneal compensated intraocular pressure (IOPcc) were measured in 56 eyes of 28 subjects without any ocular pathology, using ORA. One eye was fitted with Narafilcon A (1-Day Acuvue True Eye, Johnson & Johnson) and the other eye with Nelfilcon A (Daily AquaComfort Plus, Ciba Vision), each with −3.00D and IOPg and IOPcc were again measured over CL. The variation in the IOP with and without CL was determined. Results: Out of 28 subjects, 54% (15) were female. Mean age of the subjects was 29.4 ± 9.8 years. Both the IOPg and IOPcc when measured with CL, were found statistically significantly lower than without CL (p < 0.05). In subjects wearing Narafilcon A lens, IOPg and IOPcc were found 0.88 ± 2.04 mmHg and 1.55 ± 2.16 mmHg lower than without CL, respectively. Similarly, with Nelfilcon A lens, IOPg and IOPcc were found to be 1.03 ± 1.93 mmHg and 1.62 ± 3.12 mmHg lower, respectively. IOPcc was highly affected and underestimated by more than 3 mmHg in upto 36% of the subjects. Conclusion: Measurement of IOP over minus (−3.00D) CL with ORA is dependent upon CL properties when measured in normal IOP population. It showed lower IOP over Narafilcon A and Nelfilcon A soft CL in comparison to the pressures measured without lenses. IOPg was found less affected by CL. For the accurate measurement of IOP with ORA, CL should be removed.
TypeArticle
URIhttp://hdl.handle.net/1822/31213
DOI10.1016/j.clae.2014.07.002
ISSN1367-0484
Peer-Reviewedyes
AccessOpen access
Appears in Collections:CDF - OCV - Artigos/Papers (with refereeing)

Files in This Item:
File Description SizeFormat 
Abstract.pdfAbstract88,91 kBAdobe PDFView/Open

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID