Please use this identifier to cite or link to this item:
https://hdl.handle.net/1822/1339
Title: | An array of fabry-perot optical-channels for biological fluids analysis |
Author(s): | Minas, Graça Ribeiro, J. C. Martins, Júlio S. Wolffenbuttel, R. F. Correia, J. H. |
Keywords: | Biosystem Optical filter array Fabry-Perot |
Issue date: | 2004 |
Publisher: | Elsevier B.V. |
Journal: | Sensors and Actuators A: Physical |
Citation: | "Sensors and Actuators A : Physical". ISSN 0924-4247. 115 (2004) 362-267. |
Abstract(s): | This paper describes a biosystem (biological system) used to measure the concentration of biochemical substances in urine, serum, plasma or cerebrospinal fluid. Rather than just one channel, it comprises 16 optical-channels that enable the measurement of the concentration of 16 different biochemical substances. An array of 16 optical filters based on Fabry-Perot thin-films optical resonators has been designed. Each optical-channel is sensitive in a single wavelength with a full-width-half-maximum (FWHM) of 7 nm. The filter fabrication requires only four masks, used with different etch time. A commercially available band-pass optical filter with a band-pass wavelength in 450–650 nm is used. The biosystem requires only a white light source for illumination due the use of selective optical filters. |
Type: | Article |
URI: | https://hdl.handle.net/1822/1339 |
DOI: | 10.1016/j.sna.2004.03.077 |
ISSN: | 0924-4247 |
Peer-Reviewed: | yes |
Access: | Open access |
Appears in Collections: | DEI - Artigos em revistas internacionais |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
science1.pdf | 303,2 kB | Adobe PDF | View/Open |