Please use this identifier to cite or link to this item: https://hdl.handle.net/1822/1339

TitleAn array of fabry-perot optical-channels for biological fluids analysis
Author(s)Minas, Graça
Ribeiro, J. C.
Martins, Júlio S.
Wolffenbuttel, R. F.
Correia, J. H.
KeywordsBiosystem
Optical filter array
Fabry-Perot
Issue date2004
PublisherElsevier B.V.
JournalSensors and Actuators A: Physical
Citation"Sensors and Actuators A : Physical". ISSN 0924-4247. 115 (2004) 362-267.
Abstract(s)This paper describes a biosystem (biological system) used to measure the concentration of biochemical substances in urine, serum, plasma or cerebrospinal fluid. Rather than just one channel, it comprises 16 optical-channels that enable the measurement of the concentration of 16 different biochemical substances. An array of 16 optical filters based on Fabry-Perot thin-films optical resonators has been designed. Each optical-channel is sensitive in a single wavelength with a full-width-half-maximum (FWHM) of 7 nm. The filter fabrication requires only four masks, used with different etch time. A commercially available band-pass optical filter with a band-pass wavelength in 450–650 nm is used. The biosystem requires only a white light source for illumination due the use of selective optical filters.
TypeArticle
URIhttps://hdl.handle.net/1822/1339
DOI10.1016/j.sna.2004.03.077
ISSN0924-4247
Peer-Reviewedyes
AccessOpen access
Appears in Collections:DEI - Artigos em revistas internacionais

Files in This Item:
File Description SizeFormat 
science1.pdf303,2 kBAdobe PDFView/Open

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID